A simple and selective isocratic HPLC method for the determination of ondansetron hydrochloride has been developed. The separation is achieved by ion-pair reversed-phase chromatography on a Ultrasphere C8 column. A mixture of 0.3% sodium heptanesulphonate in phosphate buffer solution pH 3.0 and methanol (40/60, V/V) is used as mobile phase. The influence of pH, mobile phase composition and concentration of organic modifier on capacity factor of ondansetron has been investigated. The HPLC system is applied to the quantitative estimation of the active substance in bulk and pharmaceutical dosage forms.